FieldLab Desktop doesn't recognize FieldLab when connected with a USB cable

FieldLab Desktop doesn't recognize FieldLab when connected with a USB cable

The FieldLab contains built-in safety circuitry that reduces the USB signal to intrinsically safe levels.

Some newer PCs have USB port circuitry that further weakens the USB signal to a level below the required threshold for communication between the FieldLab and FieldLab Desktop. When the signal isn't strong enough, the computer can't communicate with the FieldLab.

Two ways to solve the issue:

  • Use a shorter USB cable.
    A 1-meter (or shorter) USB-A Male to USB-B Male cable preserves more of the signal than the standard 2-meter cable.

  • Connect through a powered USB hub. Plug the FieldLab into a USB hub connected to the computer, rather than directly into the computer.
    • Related Articles

    • Can't connect more than one Ralston device to FieldLab Desktop

      FieldLab Desktop software is designed and tested to allow you to connect multiple Ralston devices both wirelessly and via USB; including FieldLabs, Field Gauges, and Process Meters. On rare occasions Windows will assign the same COM port to two ...
    • Install USB Driver

      If you're having trouble getting your FieldLab connected to the PC with the provided USB cord, you may need to reinstall the Windows driver for the device. Installing drivers requires you to be a Windows administrator. 1 Open FieldLab Desktop on your ...
    • Reset the FieldLab Desktop App to Default Settings

      Learn how to reset all settings in FieldLab Desktop to the defaults, and remove all Test Modes and Data Sets from their respective libraries. 1 Select Preferences from the FieldLab Desktop menu 2 Select Reset All Settings 3 Confirm that you'd like to ...
    • Log pressure and/or temperature data to your computer with FieldLab Desktop

      This article applies to the following products: Field Gauge LC10, all models Field Gauge LC20, all models Process Meter LC30, all models FieldLab Digital Pressure Calibrator, all models. Things you will need: A PC running Windows. FieldLab Desktop ...
    • Connect one or more Ralston gauges to FieldLab Desktop App with Bluetooth

      What you will need: One or more Bluetooth-enabled Ralston Gauge. How to tell if your Field Gauge has Bluetooth. The FieldLab Desktop app installed on your Windows computer. Version 1.40.4 or later. Update FieldLab Desktop to the Latest Version. 1. ...
    • Recent Articles

    • FieldLab Desktop doesn't recognize FieldLab when connected with a USB cable

      The FieldLab contains built-in safety circuitry that reduces the USB signal to intrinsically safe levels. Some newer PCs have USB port circuitry that further weakens the USB signal to a level below the required threshold for communication between the ...
    • Find the serial number and model for your gauge

      Located on the external casing of your Field Gauge All Field Gauge LC20 and Process Meter LC30 gauges are shipped with a custom sticker that contains the following information. Model Number. Model numbers ending in B1 or W1 indicate the gauge has a ...
    • Troubleshoot USB Power Noise Causing Pressure Reading Drift on a Ralston Gauge

      Problem Unexpected pressure reading changes on an Ralston LC gauge when connected via USB. Root Cause Electrical noise is introduced through a low-quality USB hub and power supply interference. Noise from switch-mode power supplies (SMPS) can travel ...
    • Adjust Damping on your Ralston Gauge

      Damping slows the visual updates of a pressure gauge’s reading. The goal is to minimize minor fluctuations of the pressure reading on the screen caused by subtle variations in pressure that may create unwanted visual noise. Ralston gauges are set to ...
    • Register your Ralston Gauge

      Register your Ralston gauge to access calibration certificates and install firmware updates. This article applies to: FieldLab, Field Gauge LC10, LC20, and Process Meter LC30, all models. Things you will need: A PC running Windows. FieldLab Desktop ...