Replace Batteries in a Ralston Field Gauge

Replace Batteries in a Field Gauge LC20

  1. Power off the gauge.
  2. Remove the four screws from the back of the gauge with a 2mm hex key/Allen wrench or Phillips-head screwdriver. Older models shipped with hexagonal sockets, newer models ship with Phillips-head screws.


  3. Remove the back panel.
  4. Remove expired batteries, and replace with fresh AA batteries.


  5. Be sure to properly align the positive and negative ends to the proper contacts.
  6. Replace the back panel.
  7. Insert and tighten the four back panel screws to 5 in-lb (0.6 N-m) to maintain water resistance.
  8. Power on the gauge to confirm the new batteries are charged and installed properly.



    • Recent Articles

    • Max logging durations based on interval: Quick reference table

      Applies to: All Bluetooth-enabled models of Field Gauge LC20, Process Meter LC30 Ralston Field Gauge and Process Meter gauges log data to on-board memory when you start logging from the Fieldlab Mobile app. The logging interval you select determines ...
    • Add Readings to a Test

      Applies to FieldLab Mobile versions 2.0 and later. This walkthrough demonstrates how to add one pressure reading. You can add as many readings to a test as needed by repeating the steps below. 1. Connect the included Bluetooth antenna(s) to the ...
    • Delete a Reading from the Test screen

      Applies to FieldLab Mobile versions 2.0 and later 1. Tap the Reading Card 2. Tap Reading 3. Tap Delete Reading 3a. Confirm that you wish to delete.
    • Getting Started: Run Your First Test with FieldLab Mobile

      This guide walks you through connecting a gauge, logging data, and retrieving a completed test run. This walkthrough uses a single gauge, but FieldLab Mobile supports multiple gauges and readings in a single test. The following guide applies to ...
    • Update custom unit edits on your Field Gauge and Process Meter

      Applies to FieldLab Desktop app versions 2.0 and later Learn how to: Create a custom pressure unit in FieldLab Desktop When you edit and save an existing custom unit, it will need to be manually updated on any Field Gauge or Process Meter that had ...