Pressure±0.1% of Reading (20-100% of Full Scale)<br/> ±0.02% of Full Scale (0-20% of Full scale)
Vacuum (For ranges 300 psi/20 bar and below)±0.25% of Full Scale where Full Scale is -14.5 PSIG or -1 bar
TemperatureOperating: -10 to 50°C<br/> Process Media: -10 to 50°C<br/> Storage: -40 to 75 °C
Media CompatibilityGases and Liquids that are compatible with 316 Stainless Steel and Viton Seals
Relative Humidity0% to 90% (-10 to 35°C)<br/> 0% to 70% (35 to 50°C)
Ingress ProtectionIP67 (With both USB port and Charging port cover installed)
DimensionsWith no RF module: 9.3 x 5.0 x 2.5 in (21.0 x 12.7 x 6.4 cm)<br/> With RF module installed: 9.5x5.0 x2.5in(24.1x12.7x 6.4 cm)
Weight2.2 lb (1.0 kg)